Tektronix AWG610 Manual do Utilizador Página 417

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 466
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 416
Appendix B: Performance Verification
AWG610 Arbitrary Waveform Generator User Manual

3. Follow the substeps below to set the AWG610 Arbitrary Waveform
Generator controls:
a. Push UTILITY (front-panel)System (bottom)Factory Reset
(side)OK (side).
b. Load the MODE.WFM file.
Refer to Loading Files on page B–10 for file loading procedures.
c. Push Horizontal Menu (front panel)CLOCK (side).
d. Push 2, 0, 0, and M (SHIFT+7) keys in this order or used the general
purpose knob to set the internal clock frequency to 200 MHz.
4. Push the RUN button.
The LED above the RUN button is on.
5. Verify that the 10 MHz REF OUT clock signal amplitude is equal to or
larger than 1.0 V, and the clock signal period is about 100 ns.
6. Disconnect and remove the BNC cable.
Vista de página 416
1 2 ... 412 413 414 415 416 417 418 419 420 421 422 ... 465 466

Comentários a estes Manuais

Sem comentários