Tektronix 784A Manual do Utilizador Página 156

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
Vista de página 155
Performance Tests
4–80
TDS 684A, TDS 744A, & TDS 784A Service Manual
Figure 4–38: Sync Duty Cycle Test: One-Div Neg Pulse Waveform
H Turn the pulse generator PULSE DURATION variable control to
adjust the negative pulse so the oscilloscope’s CH1 – Width
measurement displays 400ns +/–10 ns.
H Turn the HORIZONTAL SCALE knob to set the oscilloscope time
base to 5ms/div.
H Turn the pulse generator PERIOD variable control to adjust the
period until the oscilloscope CH1 Period measurement reads
21.000ms –25/+50 ns. See Figure 4–39. Read note shown below.
NOTE. The pulse duration and period adjustments are critical in making this
measurement. If the pulse duration and/or the duty cycle are not stable, the
FLEXFMT function may not function. You must take care when making these
adjustments.
Vista de página 155
1 2 ... 151 152 153 154 155 156 157 158 159 160 161 ... 293 294

Comentários a estes Manuais

Sem comentários