Tektronix TDS500C Manual do Utilizador Página 95

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 148
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 94
Performance Tests
TDS 500C, TDS 600B & TDS 700C Performance Verification and Specifications
1–79
Figure 1–37: Line count accuracy correct result waveform
H Disconnect all test equipment (TSG 121) from the digitizing
oscilloscope.
H Turn off cursors by pressing CURSOR, then the main-menu
Function button, and, finally, Off from the side menu.
7. Check the Sync Duty Cycle.
a. Set up digitizing oscilloscope for Sync Duty Cycle Test:
H Press TRIGGER MENU.
H Press the Standard pop-up to select FlexFmt. Trigger Type should
already be set to Video.
H Press the main-menu Setup.
H Press the side-menu Field Rate.
H Use the keypad to set the field rate to 60.05 Hz (press 60.05, then
ENTER).
H Press the side-menu Lines.
H Use the keypad to set the field rate to 793 lines (press 793, then
ENTER).
Vista de página 94
1 2 ... 90 91 92 93 94 95 96 97 98 99 100 ... 147 148

Comentários a estes Manuais

Sem comentários