Tektronix TDS544A Manual do Utilizador Página 56

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 114
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 55
Performance Tests
Performance Verification Procedures
First, the upper and lower
limits are set so the test
waveform triggers within limits.
Second, a limit (here, the lower
limit) is increased until triggering
stops.
Figure 1-15: Measurement of Time Accuracy for Pulse and Glitch Triggering
3.
Confirm the trigger system is within time-accuracy limits for pulse-glitch or
pulse-width triggering (horizontal scale >1
s
)
:
a.
Set upper and lower limits that ensure triggering at 250 kHz:
Press the side-menu button Upper Limit. Use the keyboard to
set the upper limit to 4 s.
Press the side-menu button Lower Limit. Use the keypad to set
the lower limit to 500 ns.
b.
Display the test signal:
Set the horizontal SCALE to 5 s.
Set the output of the sine-wave generator for a 250 kHz, five-divi-
sion sine wave on screen. Set the vertical SCALE to 20 mV (the
waveform will overdrive the display).
Press SET LEVEL TO 50%.
c.
Check against limits:
Do the following subparts in the order listed.
Use the general purpose knob to
increase
Lower Limit readout
until triggering is lost.
CHECK that the Lower Limit readout is within 1 s to 3 s,
inclusive.
Use the keypad to return the Lower Limit to 500 ns and reestab-
lish triggering.
Vista de página 55
1 2 ... 51 52 53 54 55 56 57 58 59 60 61 ... 113 114

Comentários a estes Manuais

Sem comentários