Tektronix Thermometer P6249 Manual do Utilizador Página 55

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
Vista de página 54
Performance Verification
P6249 4 GHz 5X Active Probe Instruction Manual
43
11. Determine the percent error by using the formula below:
%Error + [
M1 * M2
0.4
* 1] 100%
The calculated error should be 2%.
12. Record the results on the test record.
NOTE. An unaccceptable error value may result if a low tolerance
termination is substituted for the recommended termination.
Vista de página 54
1 2 ... 50 51 52 53 54 55 56 57 58 59 60 ... 81 82

Comentários a estes Manuais

Sem comentários