Performance Verification
4–18
THS710A, THS720A, THS730A & THS720P Service Manual
This test checks the meter DC voltage accuracy.
1. Set the DC voltage source output level to 0 V.
2. Set up the TekScope instrument using the following steps:
METER VDC
3. Connect the TekScope instrument to the DC voltage source as shown below.
+
DC voltage
source
Banana cables
Check Meter DC Voltage