Performance Verification
THS710A, THS720A, THS730A & THS720P Service Manual
4–11
3. Set the TekScope instrument VOLTS/DIV to 100 mV/div.
4. Set the TekScope instrument SEC/DIV to 10 s/div.
5. Set the leveled sine wave generator frequency to 50 kHz.
6. Set the leveled sine wave generator output level so the peak-to-peak
measurement is between 599 mV and 601 mV.
7. Set the leveled sine wave generator frequency to 200 MHz if you are
checking a THS730A
, 100 MHz if you are checking a THS720A or a
THS720P, or 60 MHz if you are checking a THS710A.
8. Press MAG.
9. Set the TekScope instrument SEC/DIV to 10 ns/div.
10. Check that the peak-to-peak measurement is 425 mV.