Tektronix TDS1000- Manual do Utilizador Página 69

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 206
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 68
Application Examples
TDS1000/2000-Series Digital Oscilloscope User Manual
51
NOTE. The Positive Width measurement is available as an automatic
measurement in the Measure Menu, described on page 94.
The Positive Width measurement also displays when you select the
Single-Cycle Square opt ion in the AUTOSET Menu. Refer to page 82.
Measuring Rise Time
After measuring the pulse width, you decide that you need to check
the rise time of the pulse. Typically, you measure rise time between
the 10% and 90% levels of the waveform. To measure the rise time,
follow these steps:
1. Turn the SEC/DIV knob to display the rising edge of the
waveform.
Vista de página 68
1 2 ... 64 65 66 67 68 69 70 71 72 73 74 ... 205 206

Comentários a estes Manuais

Sem comentários