Tektronix Thermometer P7240 Manual do Utilizador Página 23

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 46
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 22
P7240 4 GHz 5X Active Probe User Manual
13
Functional Check
A functional check may be performed using the PROBE COM-
PENSATION connections on the front panel of the oscilloscope. See
Figure 3.
Figure 3: Probe functional check connections
1. Connect the probe to a TekConnect receptacle on the oscillo-
scope.
2. Set the oscilloscope to display the appropriate probe channel.
3. Using a ground lea d and SMT KlipChip, connect the probe
ground to the outer ri ng of the PROBE COMPENSATION
connector on the oscilloscope.
4. Using a standard t ip, hold the probe to the center conductor of the
PROBE COMPENSATION BNC connector on the oscilloscope.
5. Press AUTOSET, or adjust the oscilloscope to display a stable
calibration waveform. (You may need to adjust the probe offset to
display the waveform.)
Vista de página 22