Helpful Hints
Follow these helpful hints to make probing easier and noise free.
Low-inductance Grounding
Placing a ground pl ane on top of a package be ing probed c an
minimize ground lead length and inductance. See Figure 7.
Figure 7: Low-inductance grounding
Attach a small piece of copper clad on top of the package and
connect it to the package ground connection. Use the low-inductance
ground lead provided with the P7240 to keep the ground lead length
as short as possible.