Tektronix AWG610 Manual do Utilizador Página 397

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 466
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 396
Appendix B: Performance Verification
AWG610 Arbitrary Waveform Generator User Manual

3. Push UTILITY (front-panel)!System (bottom)!Factory Reset
(side)!OK (side).
4. Load the SINE.WFM file.
Refer to Loading Files on page B–10 for file loading procedures.
5. Push the RUN and CH1 output buttons.
The LEDs above the RUN button and CH1 output connector are on.
6. Do the following substeps using the n MKR function of the 497P to check
the harmonics and noise level:
a. Verify that the harmonics level of the spectrum displayed on the
spectrum analyzer from 0 Hz to 800 MHz is -40 dBc or less.
b. Verify that the noise level of the spectrum displayed on the spectrum
analyzer from 0 Hz to 800 MHz is -50 dBc or less.
7. Push the CH1 output button to turn off the CH1 LED.
8. Disconnect the spectrum analyzer.
Vista de página 396
1 2 ... 392 393 394 395 396 397 398 399 400 401 402 ... 465 466

Comentários a estes Manuais

Sem comentários