Tektronix DPO4000 Manual do Utilizador Página 264

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 585
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 263
Commands Listed in Alphabetical Order
Arguments
ON or 1 enables g
rouping of IEC harmonics.
OFF or 0 disables grouping of IEC harmonics.
Examples
POWER:HARMONICS:IEC:GROUP ING OFF disables grouping of IEC harmonics.
POWER:HARMONICS:IEC:GROUP ING? might return
:POWER:HARMONICS:IEC:GROU PING 1 indicating that the grouping is enabled
for IEC harmonics.
POWer:HARMonics:IEC:INPUTPOWer
Sets or returns the class D rated input power for IEC harmonics.
Conditions
This command requires a DPO4PWR application module.
Group
Power
Syntax
POWer:HARMonics:IEC:INPUT POWer <NR3>
POWer:HARMonics:IEC:INPUT POWer?
Arguments
<NR3> is the class D input power. Valid values ranges from 0 to 600 in
increments of 10. The unit of measure is watt.
Examples
POWER:HARMONICS:IEC:INPUT POWER 600 sets the class D input power to
600 W.
POWER:HARMONICS:IEC:INPUT POWER? might return
:POWER:HARMONICS:IEC:INPU TPOWER 100 indicating that the class D input
power is set to 100 W.
POWer:HARMonics:IEC:LINEFREQuency
Sets or returns the line frequency for the IEC standard.
Conditions
This command requires a DPO4PWR application module.
Group
Power
2-254 MSO4000 and DPO4000 Series Programmer Manual
Vista de página 263
1 2 ... 259 260 261 262 263 264 265 266 267 268 269 ... 584 585

Comentários a estes Manuais

Sem comentários