Commands Listed in Alphabetical Order
TRIGger:A:PU
LSEWidth:WIDth
Sets or returns the width setting for the pulse width trigger.
Group
Trigger
Syntax
TRIGger:A:PULSEWidth:WIDth <NR3>
TRIGger:A:PULSEWidth:WIDth ?
Related Commands
TRIGger:A:PULSEWidth:WHEn
Arguments
<NR3> species the pulse width in seconds.
Examples
TRIGGER:A:PULSEWIDTH:WIDTH 5.0E-6 sets the pulse width to 5 µs.
TRIGGER:A:PULSEWIDTH:WIDTH ? might return :TRIGGER:A:PULSEWIDTH:
WIDTH 2 .0000E-9
indicating that the pulse width is set to 2 ns.
TRIGger:A:RUNT? (Query Only)
Returns the current A runt trigger parameters.
Group
Trigger
Syntax
TRIGger:A:RUNT?
Examples
TRIGGER:A:RUNT? might return :TRIGGER:A:RUNT:SOURCE CH1;POLARIT Y
POSITIVE;WHEN OCCURS;WIDTH 4.0000E-9
.
TRIGger:A:RUNT:POLarity
Sets or returns the polarity for the runt trigger.
Group
Trigger
Syntax
TRIGger:A:RUNT:POLarity {EITher|NEGati ve|POSitive}
TRIGger:A:RUNT:POLarity?
MSO4000 and DPO4000 Series Programmer Manual 2-443