Commands Listed in Alphabetical Order
Syntax
POWer:HARMoni
cs:IEC:POWERFACtor <NR3>
POWer:HARMonics:IEC:POWER FACtor?
Arguments
<NR3> is the power factor. Valid values ranges from 0 to 1 in increments of 0.1.
Examples
POWER:HARMONICS:IEC:POWER FACTOR 1 sets the power factor to 1.0000.
POWER:HARMONICS:IEC:POWER FACTOR? might return
:POWER:HARMONICS:IEC:POWE RFACTOR 900.0000E-03 indicating
that the power factor is set to 0.9.
POWer:HARMonics:MIL:FUNDamental:CALCmethod
Sets or returns the measurement method for the MIL harmonics fundamental
current for use in calculating limits.
Conditions
This command requires a DPO4PWR application module.
Group
Power
Syntax
POWer:HARMonics:MIL:FUNDa mental:CALCmethod {MEAS|U SER}
POWer:HARMonics:MIL:FUNDa mental:CALCmethod?
Arguments
MEAS specifies that the value of the fundamental current used in calculating limits
is measured.
USER specifies that the v alue of the fundamental current used in calculated limits
is user defined.
Examples
POWER:HARMONICS:MIL:FUNDA MENTAL:CALCMETHOD USER sets the
measurement method to User.
POWER:HARMONICS:MIL:FUNDA MENTAL:CALCMETHOD? might return
:POWER:HARMONICS:MIL:FUND AMENTAL:CALCMETHOD MEAS indicating that
the measurement method is set to Meas.
POWer:HARMonics:MIL:FUNDamental:USER:CURRent
Sets or returns RMS amperes for USER CALCmethod.
Conditions
This command requires a DPO4PWR application module.
2-256 MSO4000 and DPO4000 Series Programmer Manual